Expand this Topic clickable element to expand a topic
Skip to content
Optica Publishing Group

Integrated digital holographic microscopy based on surface plasmon resonance

Open Access Open Access

Abstract

We propose a novel digital holographic microscopy (DHM) by integrating surface plasmon holographic microscopy (SPHM) with reflection DHM based on the angular and polarization multiplexing techniques. Taking advantages of the high sensitivity of surface plasmon resonance (SPR) and the high reflectivity of gold film, the tiny variations of specimen’s refractive index (RI) can be measured by using SPHM, and meanwhile, the thickness changes of the specimen can be determined by means of reflection DHM. We experimentally monitor the volatilization process of an alcohol-water mixture droplet to verify the validity of the integrated DHM. The proposed microscopy is very promising in the objective-coupling SPR microscopy for multi-information measurements of diverse specimens with low-contrast RI distributions (biomolecules, nanofluids, etc.) in a dynamic and nondestructive way.

© 2018 Optical Society of America under the terms of the OSA Open Access Publishing Agreement

1. Introduction

Surface plasmon resonance (SPR) is widely applied in the areas of physical property characterization and biochemical monitoring with the advantage of sensitive response to tiny variations of specimens in the near field of metal surface [1–4]. By measuring the intensity of the light wave reflected from the metal surface during SPR, numerous applications are experimentally achieved including nanoparticle concentration measurements [5], graphene layer characterizations [6], biomolecule dot and living cell imaging [7,8], etc. Compared with the intensity measurement, phase interrogation of the reflected light wave shows much higher sensitivity to the specimen’s refractive index (RI) change in the near field of metal surface [9]. Thus, it can be applied to measuring mixture RI [10,11], sensing protein interactions [12], monitoring DNA absorptions [13], and so on.

As an advanced imaging technique, digital holographic microscopy (DHM) can realize dynamic, wide-field, and nondestructive measurements for the complex amplitude of the light wave [14–17]. Taking these advantages, surface plasmon holographic microscopy (SPHM) is proposed to simultaneously acquire the amplitude- and phase-contrast SPR images of different specimens with high sensitivity [18,19]. Compared with total internal reflection fluorescence (TIRF) microscopy, SPHM demonstrates the superiority of non-intrusive reflection imaging with high imaging contrast. Recently, we have proposed and improved a common-path SPHM to monitor the tiny variations of the specimen RI [20,21]. We have also developed a compact objective-coupling SPHM to measure the thickness distribution of a thin film [22]. However, the current SPHM is not capable to measure the RI and thickness of a specimen simultaneously. To solve this problem, an effective approach is proposed to combine the total internal reflection DHM and transmission DHM together for determining the RI and thickness variations of the ultraviolet curing adhesive in real time [23]. Another method is demonstrated to monitor the dynamic evaporation process of a deionized water droplet from the transmission and reflection phase images, calculating the RI and thickness changes of the specimen simultaneously [24]. Regrettably, these methods above lack the high sensitivity to measure the tiny variations of the specimen RI in a dynamic process, which limits their potentials to characterize specimens with low-contrast RI distributions in the near-field detection. Consequently, the simultaneous measurements of specimen’s RI and thickness with high sensitivity require more research interests especially in the biological monitoring area.

In this paper, we firstly integrate SPHM with reflection DHM to simultaneously monitor the tiny variations of RI and thickness changes for dielectric specimens by employing the angular and polarization multiplexing techniques. Particularly, the gold film is coated on the prism in the experimental setup to simultaneously excite SPR and realize the reflection DHM.-Therefore, the tiny variations of the specimen RI can be determined with high sensitivity via the reflection phase shift of the reconstructed object waves using SPHM. While the thickness changes of specimens can be obtained from the reflection phase of the reconstructed object waves by reflection DHM. As a verification, the volatilization process of an alcohol-water droplet is monitored by simultaneously characterizing the tiny variations of RI and thickness changes of the droplet. With the advantages of dynamic, non-contact and wide-field measurements, the integrated DHM is applicable to monitoring specimens with low-contrast RI distributions especially in the biological and chemical areas.

2. Theory

2.1 Digital holographic microscopy

In DHM, the hologram is recorded in the image plane of the object, and the complex amplitude of the object wave is generally reconstructed using the angular spectrum method. In order to remove the background noise and the influence of the reference wave distortion, the double-exposure holographic interferometry is employed to obtain the relative phase distribution of the reconstructed object waves [20–24]. The relative phase distribution can be written as Δϕ(x, y) = arg[O(x, y)/O'(x, y)] = ϕO(x, y)–ϕO'(x, y), where O(x, y) and O'(x, y) denote the complex amplitudes of the reconstructed object waves with and without specimens, respectively, ϕO(x, y) and ϕO'(x, y) are the corresponding phase distributions.

2.2 RI measurement by SPHM

Figure 1 shows the three-layer Kretschmann configuration which consists of a prism, a gold layer, and a dielectric layer. In order to excite SPR, a p-polarized light beam passes through the prism at a certain incident angle (i.e. SPR angle) with a dielectric specimen adhered on the gold surface. Under the wavevector-matching condition, the light wave will resonantly couple with the surface plasmon wave (SPW) which is the electron density wave in the gold layer [25]. The penetration of SPW into the dielectric layer leads to the high sensitivity of SPR to the tiny changes of specimens in the near field of the gold surface.

 figure: Fig. 1

Fig. 1 Three-layer Kretschmann configuration. SPW: surface plasmon wave.

Download Full Size | PDF

According to the Fresnel formulas for multi-layer reflection, the complex reflection coefficient r of the light wave reflected from the gold layer is determined by multiple parameters and can be described as

r=r(ε1,ε2,ε3,d,λ,θ),
where ε1, ε2, and ε3 denote the dielectric constants of the prism, the gold layer, and the dielectric layer, respectively; d is the thickness of the gold layer; λ and θ represent the wavelength and the incident angle of the light wave, respectively. The phase shift of the reflected light wave is given by φ(x, y) = arctan{Im[r(x, y)]/Re[r(x, y)]}. Assuming that φ(x, y) and φ'(x, y) represent the phase shifts of the reflected light waves with and without specimens, respectively, the reflection phase shift difference can be expressed as Δφ(x, y) = φ(x, y)–φ'(x, y). Obviously, when a monochromatic light beam is incident on the gold layer through a specific prism with a fixed angle, the reflection phase shift difference Δφ is solely decided by the RI of the dielectric layer n3. Figure 2(a) depicts the theoretical curve between Δφ and n3, in which the quite steep slope in the blue rectangle can be used to measure the tiny variations of n3 by Δφ with high sensitivity. Because the change of reflection phase shift corresponds to the wavefront phase change of the reflected light wave, Δφ is equal to Δϕ1 which is measured by the double-exposure holographic interferometry in SPHM.

 figure: Fig. 2

Fig. 2 (a) Theoretical curve of Δφ versus n3. The thickness and dielectric constant of the gold layer are 48.8 nm and –11.740 + 1.2611i, respectively; the wavelength and incident angle of the light wave are 632.8nm and 72.82°, respectively; the RI of the prism is 1.5151. (b) 5th polynominal fitting curve of n3 versus Δφ.

Download Full Size | PDF

Since the Fresnel formulas are complex [21] and it is difficult to deduce a function between n3 and Δφ, we characterize their mathematical relationship by using fifth polynominal fitting tool in the Matlab software to calculate n3 from Δφ, as shown in Fig. 2(b) [20]. The fitting function can be expressed as

n3(x,y)=p1[Δφ(x,y)]5+p2[Δφ(x,y)]4+p3[Δφ(x,y)]3+p4[Δφ(x,y)]2+p5Δφ(x,y)+p6,
where the fitting parameters are obtained as: p1 = 8.016 × 10−12, p2 = –5.028 × 10−10, p3 = –1.125 × 10−8, p4 = 7.578 × 10−7, p5 = 3.052 × 10−5, and p6 = 1.336.

2.3 Thickness determination by reflection DHM

Owing to the high reflectivity of gold layer, a light beam which vertically hits on the gold layer is reflected back by the gold surface and goes through the specimen adhered on the gold surface twice. The relative phase distribution Δϕ2(x, y) of the reflected light wave can be determined by the double-exposure holographic interferometry in reflection DHM and can be expressed as

Δϕ2(x,y)=4π[n3(x,y)n0(x,y)]h(x,y)λ,
where h(x, y) represents the thickness of the specimen and n0(x, y) is equal to 1 when there is no specimen on the gold surface. With the prior knowledge of n3(x, y) from Eq. (2), the thickness of the specimen can be determined by

h(x,y)=Δϕ2(x,y)λ4π[n3(x,y)1].

Combining the advantages of the high sensitivity of SPR and the high reflectivity of gold film together, an integrated DHM can be constructed based on the prism-coupling SPR setup to measure the specimen’s RI and thickness simultaneously according to Eqs. (2) and (4).

3. Experimental setup

Figure 3 depicts the experimental setup of the integrated DHM. Here, we combine two off-axis Mach-Zehnder interferometers together employing the angular and polarization multiplexing techniques. In particular, a light beam from a He-Ne laser (λ = 632.8 nm) is divided into two beams by the fiber coupler FC. The two beams are expanded, collimated and polarized to 45° via the beam expanders BE1, 2, lenses L1, 2 and half-wave plates HP1, 2, respectively. One beam is further divided into p- and s-polarized ones via polarized beam splitter PBS1. The p-polarized one acts as the object beam Op in SPHM, which passes through the rectangular prism at an incident angle of 72.82° to excite SPR with a specimen adhered on the gold film. The s-polarized one acts as the object beam Os in reflection DHM, which vertically hits on the gold layer and is reflected back by the gold surface, going through the specimen twice. The two object beams further pass through the long working distance microscope objectives LWDMO1, 2 (Mitutoyo M Plan Apo 5 × ), respectively, and then are combined together through BS2 and imaged on the CCD target (Basler acA2040-90μm, 2048H × 2048V, pixel size 5.5μm) by L3. The other beam from the FC is also divided into p- and s-polarized ones via PBS2. The two divided beams act as the reference beams Rp and Rs in SPHM and reflection DHM, respectively. Finally, all the beams are combined by BS4 to form a composite hologram on the CCD target. Here, the two object beams are projected on the CCD target coaxially, while the incident angles of reference beams on the CCD target are adjusted to generate two sets of interference fringes with different orientations. The inset shows the routes of the two object beams.

 figure: Fig. 3

Fig. 3 Experimental setup of the integrated DHM. FC: fiber coupler; BE1, 2: beam expanders; L1-3: lenses; HP1, 2: half-wave plates; PBS1, 2: polarized beam splitters; M1-5: mirrors; BS1-4: beam splitters; LWDMO1, 2: long working distance microscope objectives; Op and Rp: object and reference beams in SPHM; Os and Rs: object and reference beams in reflection DHM.

Download Full Size | PDF

4. Experiment results

4.1 Preliminary measurements

To demonstrate the feasibility of the integrated DHM, we firstly measure the RI and thickness distributions of an alcohol-water mixture droplet with the volume ratio of 2:3 at the initial time. When the droplet is adhered on the gold surface, a hologram is immediately recorded. Then, a background hologram is captured after cleaning out the mixture. As shown in Fig. 4(a), there are two sets of interference fringes with nearly orthogonal directions in the composite background hologram, which effectively eliminates their crosstalk. Figure 4(b) depicts the spatial spectra of the composite hologram, where the spatial spectra in the red and yellow rectangles correspond to the object waves in SPHM and reflection DHM, respectively.

 figure: Fig. 4

Fig. 4 (a) Composite background hologram. (b) Corresponding spatial spectra.

Download Full Size | PDF

Employing the reconstruction algorithm, the relative reflection phase distribution Δϕ1(x, y) is obtained, which is equal to the reflection phase shift difference distribution Δφ(x, y). As shown in Fig. 5(a), the phase image of the specimen is compressed in one direction due to the oblique incidence of the light beam on the prism. To recover the phase image of the droplet, the reconstruction algorithm based on the angular spectrum method in digital holography is adopted to compensate for the tilted angle caused by the inclined SPR plane on the prism with respect to the optical axis [26]. Figure 5(b) shows the recovered phase image, in which all the area of the specimen image is focused for calibration. Figure 5(c) depicts the RI distribution along the black line in Fig. 5(b), according to Eq. (2). Since alcohol is completely miscible with water, the alcohol-water mixture is reasonably considered as a homogenous liquid. Thus, the RI value of the alcohol-water mixture is averaged as n3 = 1.3397 RIU. Figure 5(d) plots the reconstructed relative reflection phase distribution Δϕ2(x, y) in reflection DHM. According to Eq. (4), the three-dimensional (3D) thickness distribution of the mixture droplet is depicted in Fig. 5(e). Figure 5(f) depicts the thickness profile of the mixture droplet along the black line in Fig. 5(d).

 figure: Fig. 5

Fig. 5 Measurement results of an alcohol-water mixture droplet with the volume ratio of 2:3 at the initial time. Relative reflection phase distribution (a) Δϕ1(x, y) and (d) Δϕ2(x, y) of the specimen. (b) Undistorted phase image of (a). (c) One dimensional RI distribution of the specimen along the black line in (b). (e) 3D thickness distribution of the specimen. (f) Thickness profile of the specimen along the black line in (d).

Download Full Size | PDF

4.2 Measurements of a dynamic process

Subsequently, we monitor the volatilization process of an alcohol-water mixture droplet with the volume ratio of 1:2 to verify the dynamic measurement capability of the proposed setup. Firstly, 43 holograms with the specimen are captured at the rate of 1 fps. Then, the background hologram is recorded after removing the droplet. With numerical reconstruction, we calculate the value of relative reflection phase Δϕ1 (i.e. Δφ) by averaging the relative reflection phase distribution in the mixture area. As shown in Fig. 6(a), Δϕ1 displays a very distinct change during the volatilization process. According to Eq. (2), the corresponding RI variation of the mixture droplet is quantitatively determined. Figure 6(b) shows that the specimen RI decreases from 1.3374 RIU to 1.3318 RIU when the mixture volatilizes to pure water. So the RI variation Δn3 is 0.0056 RIU. The RI of pure water is measured as 1.3318 RIU, which is in good accordance with the measured value of 1.3317~1.3319 RIU (@20°C~25°C) by commercial devices at 632.8nm. In reflection DHM, the maximum value of relative reflection phase distribution Δϕ2 decreases continuously with the monitoring time, as depicted in Fig. 6(c). According to Eq. (4), the maximum value of the droplet thickness distribution hmax also descends gradually, as shown in Fig. 6(d). The small fluctuations of the measured values in Figs. 6(a)-6(d) result from the unstable laser source and the environmental disturbances.

 figure: Fig. 6

Fig. 6 Measurement results of monitoring the volatilization process of an alcohol-water mixture droplet with the volume ratio of 1:2. (a)-(d) Δϕ1, RI, Δϕ2 and maximum thickness variations with time of the specimen, respectively.

Download Full Size | PDF

The dynamic changes of the specimen 3D thickness distributions during the volatilization process can be seen in Visualization 1. In particular, the 3D thickness distributions of the droplet at the time of t = 0s, 9s, and 42s are shown in Figs. 7(a)–7(c), respectively. Figure 7(d) depicts the corresponding thickness profiles along the horizontal line which goes through the center of the droplet. It can be seen that the mixture droplet shrinks quite fast in the early stage, revealing the quick volatilization of alcohol. Thus, both the RI and thickness of the mixture droplet reduce successively during the monitoring time.

 figure: Fig. 7

Fig. 7 Measurement results of the 3D thickness distributions of the alcohol-water mixture droplet at (a) 0s, (b) 9s and (c) 42s, respectively. (d) Corresponding thickness profiles of the specimen (Visualization 1).

Download Full Size | PDF

5. Discussions

The specimen RI measurement based on SPHM shows very high sensitivity, which can be given by |dΔφ(x, y)/dn3(x, y)| [18]. So the RI measurement sensitivity is as high as 1.71 × 104 deg/RIU in monitoring the volatilization process of an alcohol-water mixture droplet, which is comparable to the sensitivity of 1.29 × 104 deg/RIU in the existing phase-sensitive SPR system [27]. Since the total internal reflection (TIR)-DHM can also measure the RI changes of specimens [23], we further compare the theoretical sensitivities of the specimen RI measurement based on TIR and SPR. As shown in Fig. 8, the sensitivity of the specimen RI measurement based on SPR is two orders of magnitude higher than that based on TIR, demonstrating the great advantages of monitoring both the tiny variations of specimen RI and various processes which are related to the small changes of specimen RI by using SPHM.

 figure: Fig. 8

Fig. 8 Sensitivities of the specimen RI measurement based on TIR and SPR.

Download Full Size | PDF

6. Conclusions

In conclusion, we have demonstrated an integrated DHM which can be used to simultaneously monitor the tiny variations of RI and thickness changes for dielectric specimens by combining SPHM with reflection DHM through the angular and polarization multiplexing techniques. In SPHM, the tiny variations of the specimen RI can be determined with high sensitivity based on the three-layer SPR model. The specimen thickness can be calculated from the relative reflection phase distribution in reflection DHM. The gold film coated on the prism in the experimental setup is used to simultaneously excite SPR and realize the reflection DHM. The experiment of monitoring the volatilization process of an alcohol-water mixture droplet has been successfully performed to demonstrate the effectiveness of the integrated microscopy. The sensitivity of the RI measurement is as high as 1.71 × 104 deg/RIU in the experiment. With such high sensitivity of RI measurement, our system will pave a new route to monitor the biological specimen interactions based on the objective-coupling SPR configuration in a noninvasive and dynamic way.

Funding

Key R&D Program of Shaanxi Province (2017KW-012); National Natural Science Foundation of China (NSFC) (11634010); Joint Fund of the National Natural Science Foundations of China and China Academy of Engineering Physics (NSAF) (U1730137).

Acknowledgments

We would like to thank Cheng Li at the School of Science in Northwestern Polytechnical University for the fabrication of gold film.

References

1. J. Homola, “Surface plasmon resonance sensors for detection of chemical and biological species,” Chem. Rev. 108(2), 462–493 (2008). [CrossRef]   [PubMed]  

2. L. Ji, X. Sun, G. He, Y. Liu, X. Wang, Y. Yi, C. Chen, F. Wang, and D. Zhang, “Surface plasmon resonance refractive index sensor based on ultraviolet bleached polymer waveguide,” Sens. Actuators B Chem. 244, 373–379 (2017). [CrossRef]  

3. K. S. Phillips and Q. Cheng, “Recent advances in surface plasmon resonance based techniques for bioanalysis,” Anal. Bioanal. Chem. 387(5), 1831–1840 (2007). [CrossRef]   [PubMed]  

4. H. Jussila, H. Yang, N. Granqvist, and Z. Sun, “Surface plasmon resonance for characterization of large-area atomic-layer graphene film,” Optica 3(2), 151–158 (2016). [CrossRef]  

5. I. Kim and K. D. Kihm, “Measuring near-field nanoparticle concentration profiles by correlating surface plasmon resonance reflectance with effective refractive index of nanofluids,” Opt. Lett. 35(3), 393–395 (2010). [CrossRef]   [PubMed]  

6. S. Cheon, K. D. Kihm, J. S. Park, J. S. Lee, B. J. Lee, H. Kim, and B. H. Hong, “How to optically count graphene layers,” Opt. Lett. 37(18), 3765–3767 (2012). [CrossRef]   [PubMed]  

7. G. Steiner, “Surface plasmon resonance imaging,” Anal. Bioanal. Chem. 379(3), 328–331 (2004). [CrossRef]   [PubMed]  

8. Y. Yanase, T. Hiragun, S. Kaneko, H. J. Gould, M. W. Greaves, and M. Hide, “Detection of refractive index changes in individual living cells by means of surface plasmon resonance imaging,” Biosens. Bioelectron. 26(2), 674–681 (2010). [CrossRef]   [PubMed]  

9. A. V. Kabashin, S. Patskovsky, and A. N. Grigorenko, “Phase and amplitude sensitivities in surface plasmon resonance bio and chemical sensing,” Opt. Express 17(23), 21191–21204 (2009). [CrossRef]   [PubMed]  

10. J. Zhang, Q. Dai, and G. P. Wang, “Surface plasmon interferometric microscopy for three-dimensional imaging of dynamic processes,” Opt. Lett. 31(20), 3004–3006 (2006). [CrossRef]   [PubMed]  

11. Y. H. Huang, H. P. Ho, S. Y. Wu, S. K. Kong, W. W. Wong, and P. Shum, “Phase sensitive SPR sensor for wide dynamic range detection,” Opt. Lett. 36(20), 4092–4094 (2011). [CrossRef]   [PubMed]  

12. J. Y. Lee, T. K. Chou, and H. C. Shih, “Polarization-interferometric surface-plasmon-resonance imaging system,” Opt. Lett. 33(5), 434–436 (2008). [CrossRef]   [PubMed]  

13. A. R. Halpern, Y. Chen, R. M. Corn, and D. Kim, “Surface plasmon resonance phase imaging measurements of patterned monolayers and DNA adsorption onto microarrays,” Anal. Chem. 83(7), 2801–2806 (2011). [CrossRef]   [PubMed]  

14. M. K. Kim, Digital Holographic Microscopy (Springer, 2011), Chap. 11.

15. P. Marquet, B. Rappaz, P. J. Magistretti, E. Cuche, Y. Emery, T. Colomb, and C. Depeursinge, “Digital holographic microscopy: a noninvasive contrast imaging technique allowing quantitative visualization of living cells with subwavelength axial accuracy,” Opt. Lett. 30(5), 468–470 (2005). [CrossRef]   [PubMed]  

16. W. Sun, J. Zhao, J. Di, Q. Wang, and L. Wang, “Real-time visualization of Karman vortex street in water flow field by using digital holography,” Opt. Express 17(22), 20342–20348 (2009). [CrossRef]   [PubMed]  

17. J. Wang, J. Zhao, C. Qin, J. Di, A. Rauf, and H. Jiang, “Digital holographic interferometry based on wavelength and angular multiplexing for measuring the ternary diffusion,” Opt. Lett. 37(7), 1211–1213 (2012). [CrossRef]   [PubMed]  

18. S. Li and J. Zhong, “Simultaneous amplitude-contrast and phase-contrast surface plasmon resonance imaging by use of digital holography,” Biomed. Opt. Express 3(12), 3190–3202 (2012). [CrossRef]   [PubMed]  

19. C. Hu, J. Zhong, and J. Weng, “Digital holographic microscopy by use of surface plasmon resonance for imaging of cell membranes,” J. Biomed. Opt. 15(5), 056015 (2010). [CrossRef]   [PubMed]  

20. J. Zhang, S. Dai, C. Ma, J. Di, and J. Zhao, “Common-path digital holographic microscopy for near-field phase imaging based on surface plasmon resonance,” Appl. Opt. 56(11), 3223–3228 (2017). [CrossRef]   [PubMed]  

21. J. Zhang, S. Dai, J. Zhong, T. Xi, C. Ma, Y. Li, J. Di, and J. Zhao, “Wavelength-multiplexing surface plasmon holographic microscopy,” Opt. Express 26(10), 13549–13560 (2018). [CrossRef]   [PubMed]  

22. J. Zhang, S. Dai, C. Ma, J. Di, and J. Zhao, “Compact surface plasmon holographic microscopy for near-field film mapping,” Opt. Lett. 42(17), 3462–3465 (2017). [CrossRef]   [PubMed]  

23. J. Zhang, C. Ma, S. Dai, J. Di, Y. Li, T. Xi, and J. Zhao, “Transmission and total internal reflection integrated digital holographic microscopy,” Opt. Lett. 41(16), 3844–3847 (2016). [CrossRef]   [PubMed]  

24. C. Ma, J. Di, J. Zhang, Y. Li, T. Xi, E. Li, and J. Zhao, “Simultaneous measurement of refractive index distribution and topography by integrated transmission and reflection digital holographic microscopy,” Appl. Opt. 55(33), 9435–9439 (2016). [CrossRef]   [PubMed]  

25. H. Lu, X. Gan, D. Mao, B. Jia, and J. Zhao, “Flexibly tunable high-quality-factor induced transparency in plasmonic systems,” Sci. Rep. 8(1), 1558 (2018). [CrossRef]   [PubMed]  

26. W. M. Ash 3rd, L. Krzewina, and M. K. Kim, “Quantitative imaging of cellular adhesion by total internal reflection holographic microscopy,” Appl. Opt. 48(34), H144–H152 (2009). [CrossRef]   [PubMed]  

27. W. Yuan, H. P. Ho, C. L. Wong, S. K. Kong, and C. Lin, “Surface plasmon resonance biosensor incorporated in a Michelson interferometer with enhanced sensitivity,” IEEE Sens. J. 7(1), 70–73 (2007). [CrossRef]  

Supplementary Material (1)

NameDescription
Visualization 1       The dynamic changes of the 3D thickness distributions of an alcohol-water mixture droplet during the volatilization process.

Cited By

Optica participates in Crossref's Cited-By Linking service. Citing articles from Optica Publishing Group journals and other participating publishers are listed here.

Alert me when this article is cited.


Figures (8)

Fig. 1
Fig. 1 Three-layer Kretschmann configuration. SPW: surface plasmon wave.
Fig. 2
Fig. 2 (a) Theoretical curve of Δφ versus n3. The thickness and dielectric constant of the gold layer are 48.8 nm and –11.740 + 1.2611i, respectively; the wavelength and incident angle of the light wave are 632.8nm and 72.82°, respectively; the RI of the prism is 1.5151. (b) 5th polynominal fitting curve of n3 versus Δφ.
Fig. 3
Fig. 3 Experimental setup of the integrated DHM. FC: fiber coupler; BE1, 2: beam expanders; L1-3: lenses; HP1, 2: half-wave plates; PBS1, 2: polarized beam splitters; M1-5: mirrors; BS1-4: beam splitters; LWDMO1, 2: long working distance microscope objectives; Op and Rp: object and reference beams in SPHM; Os and Rs: object and reference beams in reflection DHM.
Fig. 4
Fig. 4 (a) Composite background hologram. (b) Corresponding spatial spectra.
Fig. 5
Fig. 5 Measurement results of an alcohol-water mixture droplet with the volume ratio of 2:3 at the initial time. Relative reflection phase distribution (a) Δϕ1(x, y) and (d) Δϕ2(x, y) of the specimen. (b) Undistorted phase image of (a). (c) One dimensional RI distribution of the specimen along the black line in (b). (e) 3D thickness distribution of the specimen. (f) Thickness profile of the specimen along the black line in (d).
Fig. 6
Fig. 6 Measurement results of monitoring the volatilization process of an alcohol-water mixture droplet with the volume ratio of 1:2. (a)-(d) Δϕ1, RI, Δϕ2 and maximum thickness variations with time of the specimen, respectively.
Fig. 7
Fig. 7 Measurement results of the 3D thickness distributions of the alcohol-water mixture droplet at (a) 0s, (b) 9s and (c) 42s, respectively. (d) Corresponding thickness profiles of the specimen (Visualization 1).
Fig. 8
Fig. 8 Sensitivities of the specimen RI measurement based on TIR and SPR.

Equations (4)

Equations on this page are rendered with MathJax. Learn more.

r=r( ε 1 , ε 2 , ε 3 ,d,λ,θ),
n 3 (x,y)= p 1 [Δφ(x,y)] 5 + p 2 [Δφ(x,y)] 4 + p 3 [Δφ(x,y)] 3 + p 4 [Δφ(x,y)] 2 + p 5 Δφ(x,y)+ p 6 ,
Δ ϕ 2 (x,y)= 4π[ n 3 (x,y) n 0 (x,y)]h(x,y) λ ,
h(x,y)= Δ ϕ 2 (x,y)λ 4π[ n 3 (x,y)1] .
Select as filters


Select Topics Cancel
© Copyright 2024 | Optica Publishing Group. All rights reserved, including rights for text and data mining and training of artificial technologies or similar technologies.