Expand this Topic clickable element to expand a topic
Skip to content
Optica Publishing Group

Highly stable wide-field common path digital holographic microscope based on a Fresnel biprism interferometer

Open Access Open Access

Abstract

Quantitative phase imaging (QPI) of biological cells and tissues is an important technique useful in the determination of many biophysical parameters such as morphology, refractive index, thickness, cell dry mass, hemoglobin concentration, etc. Off-axis digital holography has been ideal for the QPI of microscopic specimens, but it has lower temporal stability compared to on-axis and common path digital holographic microscopes, which offer higher temporal stability. In this paper, we present a very simple, easy to align yet highly stable common path digital holographic microscope based on a Fresnel biprism interferometer. The system uses a biprism to divide the incoming beam into reference and object beams without the loss of optical power, unlike diffraction phase microscopy. Two methods are proposed, one by utilizing the spatial filtering mode and the other by the self-referencing mode. It offers the advantage that the reference beam can be easily created simply by translating the object in the focal plane of the microscopic objective or by spatially filtering one of the object beams in the Fourier domain. The proposed setup offers no power loss and a high phase stability of approximately 0.006 rad (6 mrad) without using any vibration isolation. The experiments on industrial and biological samples are reported demonstrating its application both for static and dynamic samples.

© 2018 Optical Society of America under the terms of the OSA Open Access Publishing Agreement

1. Introduction

Most of the biological specimens are transparent in nature thereby making their imaging difficult under a bright field microscope. Digital holographic microscopy (DHM) has been developed as a powerful and accurate measurement technique, which provides us with amplitude and phase information of the transparent samples simultaneously [1–4]. The technique is advantageous due to its non-invasive, high resolution and wide field of view capability [5]. DHM can be implemented in two modes off-axis and in-line configurations. Off-axis DHM can be achieved either by Michelson or Mach-Zehnder type setups. The off axis DHM geometry initiates independent paths for the reference and object beams and passing through separate set of optical elements. Because of two widely separated paths this configuration becomes very sensitive to environmental vibrations, fluctuations, generate large phase noise thus resulting in low spatial and temporal stability [6,7]. The photon shot noise is often the limiting factor in interferometric microscopy measurements and by using state-of-the-art CCD camera with million-level electrons full well capacity can significantly reduce shot noise [8]. But cooled CCD cameras are expensive. Further off-axis DHM is not suitable for QPI of dynamic samples.

In recent years common path interferometers has been widely used due to its robustness and high stability [2,8]. In common path interferometers both the reference and object beams travel approximately the same path thus making them less sensitive to environmental vibrations and noise. Although being highly stable the common path interferometers deals with the shortcoming of generating twin images of the specimen. This shortcoming can be overcome depending on the method of generation of the reference beam. The first method involves the usage of spatial filtering in order to generate reference beam from the two object beams [6,7,9–11]. In [6] the author has employed a pinhole in the Fourier plane in combination to τ interferometer, which is a Michelson configuration. In [12,13] the authors have employed grating to generate the two beams and spatially filtered in the Fourier domain to obtain reference beam from one of the object beams. The use of grating leads to a great deal of power loss as only 0th and 1st diffraction orders are utilized [14–16]. In [16] a diffraction grating is utilized to generate a reference beam that traverses a blank region of the sample in a common-path off-axis interferometry setup to obtain tomographic imaging. The other method of overcoming twin image problem is to use self-referencing technique [17–21]. In [20,21] the authors have employed a single beam splitter to split and combine a spherical wavefront while placing sample in half of the beam. In [7,18,22] the authors have employed a glass plate to generate two laterally sheared beams and spatial filtering of one of the beams to generate reference beam but disadvantage being that glass plate reflects only 4% of the incoming power leading to surplus optical power loss which is lost in transmission as only reflected light is used.

In this paper, we report the development of a highly stable common path DHM, which has the capability of solving the twin image problem by both methods, mentioned earlier. The proposed setup makes use of a Fresnel biprism to generate two beams by dividing the incoming beam into two equal halves. The two generated beams are then interfered with the help of an asymmetric 4f imaging system in order to achieve sufficient fringe density on the image plane. The interference of two beams can be easily obtained using the present set-up without any hassle in the alignment of the two beams, as it is the simplest method of generating interference. The two beams have equal intensity, which is an essential requirement for interferometry. Further, the overlapping image problem in the present setup can be easily solved using spatial filtering in the Fourier domain. A much simpler and easy solution to the overlapping problem can also be done by self-referencing method. This can be easily achieved in our proposed set up as the Fresnel biprism divides the original beams into equal halves through the center of the beam without compensating on power such that one half contains object information while the other half does not. The advantage of the proposed setup is that it is easy to align, no power loss, high temporal stability due to its common path nature and can be used to image dense samples. Only the proposed setup can operate in any of the two possibilities at the same time, which is not possible with any other previously reported common path techniques.

2. Experimental setup

The schematic view of the experimental setup is shown in Fig. 1

 figure: Fig. 1

Fig. 1 Experimental setup of proposed common path DHM using Fresnel biprism. OBJ is objective lens and L1, L2 and L3 are lenses of focal length 100mm, 100mm and 175mm.

Download Full Size | PDF

. The highly stable common path DHM couples light from a 532nm laser with coherence length 144 μm onto a single mode fiber and a collimator. The incoming beam illuminates the sample mounted on a translational stage and the light scattered from the sample is collected by a 20X microscopic objective lens and collimated by the lens L1. The collected beam is made to fall on a Fresnel biprism, which divides the beam into two equal parts and makes them travel at an angle with one another. One beam being the reference beam and the other being object beam. The beams are interfered on the image plane using an asymmetric 4f imaging lens system (L2 and L3, 100mm and 175 mm focal length respectively). The asymmetric lens system offers an additional magnification of 1.75X which helps in achieving sufficient fringe density on the image plane such that it satisfies Nyquist sampling criteria. The two beams interfere in the overlapping region and are recorded using a CCD camera (ARTRAY PIII). The beams are approximately 4mm apart at the focal plane of L2, which can be easily varied depending on focal length of L2. Both the interfering beams carry object information thus the final image faces the problem of overlapping.

3. Results and discussion

The hologram recorded by CCD camera is processed using Fourier fringe analysis technique (Fig. 2

 figure: Fig. 2

Fig. 2 (a) Hologram recorded having overlapping problem and (b) reconstructed amplitude.

Download Full Size | PDF

). The recorded hologram was Fourier transformed and its 1st order peak was filtered out using a circular filter. Then amplitude and phase information was extracted from the image after taking inverse Fourier transform. A reference hologram was also recorded for each of the object samples for phase compensation and finally the phase was calculated by subtracting this background phase. The phase difference has been calculated using the formulae [7]
Δϕ=2π(ncnr)t/λ
where nc and nr are the refractive indices of the object and the surrounding medium, t is the thickness of the medium and λ is the wavelength used. Here, (nc - nr)t corresponds to the optical thickness of the specimen.

The overlapping issue in our case can be easily resolved using spatial filtering or self-referencing method. In Fig. 3

 figure: Fig. 3

Fig. 3 Overcoming the overlapping problem using (a) spatial filtering (inset 1) and (b) translating the sample (inset 2).

Download Full Size | PDF

, inset 1 shows the method of solving the overlapping image problem using spatial filtering in the Fourier plane. Solution is achieved by inserting a 25 µm pinhole in the path of one of the beams without blocking the other. The pinhole passes only the DC component of the beam and higher frequencies are cut off such that it acts as reference beam and the interference pattern recorded by the CCD camera becomes free from overlapping problem. The method is little tedious and time taking as it involves passing only one of the beams through a 25 µm pinhole without blocking the other as the beams are only 4mm apart. In our proposed set up a much simpler and easy solution to the overlapping problem can be accomplished just by translating the sample in the focal plane of the objective lens such that it covers only one half of the field of view of the objective lens while the other half acts as reference beam. Thus, the method works by subdividing the beam into half object beam and half reference beam such that the object is located only in the object portion of the beam while other half passes unaltered as depicted in inset 2. The advantage of the proposed method is that it makes it extremely easy to align and there is no loss of power, which takes place while using grating or beam splitters. Further the system offers high temporal stability and can be used for imaging dense samples.

Figure 4

 figure: Fig. 4

Fig. 4 (a) Hologram recorded of USAF 1951 resolution chart with spatial filtering and (b) reconstructed amplitude.

Download Full Size | PDF

shows the hologram recorded when a 25 µm pinhole was inserted in the path of one of the beams such that it stops the object information and acts as reference beam. Figure 4(a) shows the hologram recorded for the resolution chart while 4b shows the reconstructed amplitude image. As compared to Fig. 2(a) and 2(b) the overlapping in hologram and in the reconstructed images are completely removed in Fig. 4(a), 4(b). The experiment was also carried out on human erythrocytes (RBC) as shown in Fig. 5
 figure: Fig. 5

Fig. 5 (a) Hologram recorded of RBC sample with spatial filtering. (b) Zoomed in image of hologram and (c) reconstructed phase map.

Download Full Size | PDF

. Figure 5(b) shows the zoomed in image of the hologram. The phase distribution for the corresponding RBCs is shown in Fig. 5(c).

The system proposed has high temporal stability due to its common path nature. The temporal stability of the set up was measured by recording a series of holograms with blank microscopic slide as sample. All the experiments have been performed without any vibration isolation. The hologram recording was done at the frame rate of 10 frames/sec for 15 secs. The phase difference was calculated for each of the frame initializing first frame as the reference frame. Standard deviation of the phase difference was measured for 10^4 random pixel locations giving us the mean standard deviation of 0.006 rad. Figure 6(a)

 figure: Fig. 6

Fig. 6 Temporal stability of proposed setup. (a) Histogram of the standard deviation of phase difference of proposed setup with a mean variation of 0.006 rad. (b) Histogram of the standard deviation of phase difference of Mach-Zehnder setup with a mean variation of 0.2 rad.

Download Full Size | PDF

shows the histogram of the standard deviation values with a mean variation as 0.006 rad. On comparing with the Mach Zehnder setup by means of keeping the environmental conditions exactly same, i.e. without any vibration isolation, gave the mean variation of phase 0.2 rad [as shown in Fig. 6(b)], which is similar to the reported values earlier [2]. Therefore, the temporal stability of the proposed set up has significantly improved compared to the off-axis configurations.

In order to study the temporal versatility of our experimental setup for the measurement of dynamic samples we have studied the evaporation of acetone (Fisher scientific, (CH3)2CO Reagent grade) with time. The holograms were recorded using CCD camera at a frame rate of 10 frames/sec for a total of 4 seconds until the reagent completely evaporates. The refractive index of acetone as measured with Abbe refractometer (1.361) was used to calculate height of the droplets. The variation of height profile of droplets of acetone as it evaporates when exposed to air has been shown in Fig. 7

 figure: Fig. 7

Fig. 7 Evaporation of acetone (CH3)2CO with time. 3D height profile of acetone at time instants 0.24s, 1.52s, 2s, 2.48s, 2.56s, and 3.12s. Color bar represents height in micrometers.

Download Full Size | PDF

at the time instants of 0.24s, 1.52s, 2.00s, 2.48s, 2.56 and 3.12s. A background phase was measured for the glass slide without the sample and was subtracted from all recorded holograms to minimize the background noise. Figure 7 shows the evaporation process as the height of acetone droplets decreases from 1µm to 0µm.

As discussed earlier that though introduction of pinhole in one of the beams overcomes the drawback of overlapping object but it is a bit difficult to align and tedious. A much simpler way is to translate the sample such that it occupies only half of the field of view of the objective lens while the other half acts as reference beam. This method has been described in Fig. 3 inset 2. To demonstrate this, the initial experiment was carried on a USAF 1951 resolution chart where the resolution chart lies in the right beam and the left beam acts as reference beam. Figure 8(a)

 figure: Fig. 8

Fig. 8 (a) Hologram recorded of USAF 1951 resolution chart by translating the sample and (b) reconstructed amplitude.

Download Full Size | PDF

shows the recorded hologram and Fig. 8(b) shows the reconstructed amplitude image without the overlapping problem achieved using self-reference mode. The experiment was also carried out on human RBC as shown in Fig. 9
 figure: Fig. 9

Fig. 9 (a) Hologram recorded of RBC sample by translating the sample. (b) Zoomed in image of hologram (c) reconstructed phase image of red square in (b).

Download Full Size | PDF

. Figure 9(a) shows the hologram recorded for the RBC with the help of CCD camera. Here only right beam contains the object information while the left beam acts as the reference beam. This makes the operation of the proposed DHM set up very easy. Figure 9(b) shows zoomed in image of the hologram and Fig. 9(c) shows the reconstructed phase profile of RBC. The area of the sample that was measured in the setup was 0.10 mm2 and the maximum distance from the edge of the sample was 0.33mm.

4. Conclusion

We have demonstrated a very easy to implement and align common path DHM using a Fresnel biprism. The object and reference beams are generated using spatial filtering and self-referencing methods. The holographic set-up is the simplest to align among all the common path configurations reported so far. The two beams that pass through the same set of optical elements and interfere in the camera plane to produce hologram with a temporal stability of 0.006 rad without any vibration isolation. The temporal stability of the present set-up is much better than the off-axis configuration. Therefore, the proposed DHM is most suitable for QPI of dynamic samples such as cell membrane fluctuations and QPI of micro-fluidic channels. Further, the setup is compact and robust which can be easily deployed for field studies without any vibration isolation. Experiments with industrial and biological samples have been shown in the two proposed arrangements namely by use of spatial filtering and translation of the sample. Though it suffers with disadvantage of offering only half field of view, offers the advantage of being simple, amenable, high temporal stability with no power loss.

Funding

Department of Science and Technology (IDP/MED/14/2016/General).

Acknowledgments

The authors are thankful to the Department of Science and Technology for funding the project no. IDP/MED/14/2016/General.

References

1. M. K. Kim, “Principles and techniques of digital holographic microscopy,” SPIE Reviews 1(1), 018005 (2010).

2. C. Ma, Y. Li, J. Zhang, P. Li, T. Xi, J. Di, and J. Zhao, “Lateral shearing common-path digital holographic microscopy based on a slightly trapezoid Sagnac interferometer,” Opt. Express 25(12), 13659–13667 (2017). [CrossRef]   [PubMed]  

3. B. Kemper, A. Vollmer, C. E. Rommel, J. Schnekenburger, and G. von Bally, “Simplified approach for quantitative digital holographic phase contrast imaging of living cells,” J. Biomed. Opt. 16(2), 026014 (2011). [CrossRef]   [PubMed]  

4. K. Lee and Y. Park, “Quantitative phase imaging unit,” Opt. Lett. 39(12), 3630–3633 (2014). [CrossRef]   [PubMed]  

5. J. Zheng, P. Gao, and X. Shao, “Opposite-view digital holographic microscopy with autofocusing capability,” Sci. Rep. 7(1), 4255 (2017). [CrossRef]   [PubMed]  

6. N. T. Shaked, “Quantitative phase microscopy of biological samples using a portable interferometer,” Opt. Lett. 37(11), 2016–2018 (2012). [CrossRef]   [PubMed]  

7. H. Bai, M. Shan, Z. Zhong, L. Guo, and Y. Zhang, “Common path interferometer based on the modified Michelson configuration using a reflective grating,” Opt. Lasers Eng. 75, 1–4 (2015). [CrossRef]  

8. P. Hosseini, R. Zhou, Y. H. Kim, C. Peres, A. Diaspro, C. Kuang, Z. Yaqoob, and P. T. C. So, “Pushing phase and amplitude sensitivity limits in interferometric microscopy,” Opt. Lett. 41(7), 1656–1659 (2016). [CrossRef]   [PubMed]  

9. P. Vora, V. Trivedi, S. Mahajan, N. Patel, M. Joglekar, V. Chhaniwal, A. R. Moradi, B. Javidi, and A. Anand, “Wide field of view common-path lateral-shearing digital holographic interference microscope,” J. Biomed. Opt. 22(12), 1–11 (2017). [PubMed]  

10. H. Bai, M. Shan, Z. Zhong, L. Guo, and Y. Zhang, “Parallel-quadrature on-axis phase-shifting common-path interferometer using a polarizing beam splitter,” Appl. Opt. 54(32), 9513–9517 (2015). [CrossRef]   [PubMed]  

11. S. Mahajan, V. Trivedi, P. Vora, V. Chhaniwal, B. Javidi, and A. Anand, “Highly stable digital holographic microscope using Sagnac interferometer,” Opt. Lett. 40(16), 3743–3746 (2015). [CrossRef]   [PubMed]  

12. M. Shan, B. Hao, Z. Zhong, M. Diao, and Y. Zhang, “Parallel two-step spatial carrier phase-shifting common-path interferometer with a Ronchi grating outside the Fourier plane,” Opt. Express 21(2), 2126–2132 (2013). [CrossRef]   [PubMed]  

13. Y. Kim, H. Shim, K. Kim, H. Park, J. H. Heo, J. Yoon, C. Choi, S. Jang, and Y. Park, “Common-path diffraction optical tomography for investigation of three-dimensional structures and dynamics of biological cells,” Opt. Express 22(9), 10398–10407 (2014). [CrossRef]   [PubMed]  

14. B. Bhaduri, C. Edwards, H. Pham, R. Zhou, T. H. Nguyen, L. L. Goddard, and G. Popescu, “Diffraction phase microscopy: principles and applications in materials and life sciences,” Adv. Opt. Photonics 6(1), 57–119 (2014). [CrossRef]  

15. G. Popescu, T. Ikeda, R. R. Dasari, and M. S. Feld, “Diffraction phase microscopy for quantifying cell structure and dynamics,” Opt. Lett. 31(6), 775–777 (2006). [CrossRef]   [PubMed]  

16. W. C. Hsu, J. W. Su, T. Y. Tseng, and K. B. Sung, “Tomographic diffractive microscopy of living cells based on a common-path configuration,” Opt. Lett. 39(7), 2210–2213 (2014). [CrossRef]   [PubMed]  

17. J. Jang, C. Y. Bae, J. K. Park, and J. C. Ye, “Self-reference quantitative phase microscopy for microfluidic devices,” Opt. Lett. 35(4), 514–516 (2010). [CrossRef]   [PubMed]  

18. A. Anand, V. Chhaniwal, S. Mahajan, V. Trivedi, A. Singh, R. Leitgeb, and B. Javidi, “Self-referencing digital holographic microscope for dynamic imaging of living cells,” Proc. SPIE 9117, 91170X (2014). [CrossRef]  

19. P. Bon, G. Maucort, B. Wattellier, and S. Monneret, “Quadriwave lateral shearing interferometry for quantitative phase microscopy of living cells,” Opt. Express 17(15), 13080–13094 (2009). [CrossRef]   [PubMed]  

20. W. Qu, K. Bhattacharya, C. O. Choo, Y. Yu, and A. Asundi, “Transmission digital holographic microscopy based on a beam-splitter cube interferometer,” Appl. Opt. 48(15), 2778–2783 (2009). [CrossRef]   [PubMed]  

21. W. Qu, O. C. Chee, Y. Yu, and A. Asundi, “Characterization and inspection of microlens array by single cube beam splitter microscopy,” Appl. Opt. 50(6), 886–890 (2011). [CrossRef]   [PubMed]  

22. A. S. G. Singh, A. Anand, R. A. Leitgeb, and B. Javidi, “Lateral shearing digital holographic imaging of small biological specimens,” Opt. Express 20(21), 23617–23622 (2012). [CrossRef]   [PubMed]  

Cited By

Optica participates in Crossref's Cited-By Linking service. Citing articles from Optica Publishing Group journals and other participating publishers are listed here.

Alert me when this article is cited.


Figures (9)

Fig. 1
Fig. 1 Experimental setup of proposed common path DHM using Fresnel biprism. OBJ is objective lens and L1, L2 and L3 are lenses of focal length 100mm, 100mm and 175mm.
Fig. 2
Fig. 2 (a) Hologram recorded having overlapping problem and (b) reconstructed amplitude.
Fig. 3
Fig. 3 Overcoming the overlapping problem using (a) spatial filtering (inset 1) and (b) translating the sample (inset 2).
Fig. 4
Fig. 4 (a) Hologram recorded of USAF 1951 resolution chart with spatial filtering and (b) reconstructed amplitude.
Fig. 5
Fig. 5 (a) Hologram recorded of RBC sample with spatial filtering. (b) Zoomed in image of hologram and (c) reconstructed phase map.
Fig. 6
Fig. 6 Temporal stability of proposed setup. (a) Histogram of the standard deviation of phase difference of proposed setup with a mean variation of 0.006 rad. (b) Histogram of the standard deviation of phase difference of Mach-Zehnder setup with a mean variation of 0.2 rad.
Fig. 7
Fig. 7 Evaporation of acetone (CH3)2CO with time. 3D height profile of acetone at time instants 0.24s, 1.52s, 2s, 2.48s, 2.56s, and 3.12s. Color bar represents height in micrometers.
Fig. 8
Fig. 8 (a) Hologram recorded of USAF 1951 resolution chart by translating the sample and (b) reconstructed amplitude.
Fig. 9
Fig. 9 (a) Hologram recorded of RBC sample by translating the sample. (b) Zoomed in image of hologram (c) reconstructed phase image of red square in (b).

Equations (1)

Equations on this page are rendered with MathJax. Learn more.

Δϕ=2π( n c n r )t/λ
Select as filters


Select Topics Cancel
© Copyright 2024 | Optica Publishing Group. All rights reserved, including rights for text and data mining and training of artificial technologies or similar technologies.