Abstract
Triple phase grating X-ray interferometry is a promising new technique of grating based X-ray differential phase contrast imaging. Accurate retrieval of sample phase gradients from measured interference fringe shifts is a key task in X-ray interferometry. To fulfill this task in triple phase grating X-ray interferometry with monochromatic X-ray sources, the authors derived exact formulas relating sample phase gradient to fringe phase shift. These formulas not only provide a design optimization tool for triple phase grating interferometry, but also lay a foundation for quantitative phase contrast imaging.
© 2020 Optical Society of America under the terms of the OSA Open Access Publishing Agreement
1. Introduction
X-ray phase contrast imaging, which holds promising potential for medical imaging, can also be implemented by grating-based X-ray interferometry [1–16]. Currently most of the grating-based X-ray interferometry is implemented as Talbot-Lau interferometry, in which a single phase grating is employed as a beam splitter to split the X-ray into diffraction orders. The interference between the diffracted orders forms intensity fringes. High-contrast interference fringes can be formed at certain so-called Talbot-distances downstream [1–4]. To increase the grating interferometer’s sensitivity, fine-pitch phase gratings of periods as small as a few micrometers or even sub-micrometers should be used. In this way the intensity fringe period is of comparable size as that of the phase grating period. Nevertheless, in medical imaging, it is only feasible to utilize hospital-grade imaging detectors, whose pixels are of a few tens of micrometers. To enable the fringe detection with common image detectors, one method is to use a fine-pitch absorbing grating placed at the entrance of the detector. This analyzer grating enables one to indirectly detect the fringe patterns through grating scanning [1–4]. By analyzing the fringes, a sample attenuation image, a phase gradient image, and a dark field image are reconstructed. However, the absorbing grating analyzer blocks more than half of the information-carrying X-ray, thereby greatly increases radiation dose in the imaging studies. This is a disadvantage of the Talbot-Lau X-ray interferometry for medical imaging applications, where radiation dose involved should be as low as reasonably achievable.
The recent emergence of dual and triple phase grating interferometry brings a solution for the dose inefficiency problems of the Talbot-Lau interferometry [17,18]. One of the unique advantages with the use of dual and triple phase gratings is that it enables the use of hospital-grade imaging detectors for resolving intensity fringes generated by the fine-pitch phase gratings, without the need of an absorbing grating [17,18]. In a previous work [19–21], based on a quantitative theory of the dual phase grating interferometry, we clarified the mechanism of the intensity fringe formation in dual phase grating interferometry. Indeed, each phase grating splits impinging X-ray into different diffraction orders. The diffracted waves interfere with one another, creating intensity fringes of different diffracted orders. A common imaging detector, through its pixel averaging effects, resolves only the beat patterns of large periodicities. Ge et al. [22] analyzed the formation mechanism of the dual-phase grating interferometry from the geometrical optics point of view. In this work, the understanding gained from the dual phase grating interferometry is extended to the triple phase grating interferometry. We found that the same fringe formation mechanism works for triple phase grating interferometry, as will be explained in Section 2. A rigorous analysis for triple phase grating interferometry is given in Appendix.
The focus of this work is to understand, for triple phase grating interferometry, what are the attributes of an interferometry setup that determine the fringe phase shift ensued from sample refraction. This is an important task for interferometry for the following reasons. When a sample is introduced in an interferometer, sample refracts X-ray and distorts the original intensity fringe. Specifically, the local refraction angle $\alpha (x,y)$ is proportional to the sample’s phase gradient $\partial \Phi _{\mathrm {s}} (x,y)/\partial x$[3,4]:
where $\lambda$ is the X-ray wavelength, and $\Phi _{\mathrm {s}}(x,y)$ is the sample’s X-ray phase shift. Of note, $\Phi _{\mathrm {s}}(x,y) = -\lambda r_{e}\int \rho _{e}(x,y,z)\,\mathrm {d} z$, where $\rho _{e}(x,y,z)$ is the sample electron density distribution and $r_e$ is the classical electron density which equals to $2.82\mathrm {E}\!-\!\!15$m. One important kind of fringe distortion is the localized fringe phase shifts $\Delta \phi (x,y)$, which can be measured through phase stepping or Fourier analysis of intensity fringe pattern [1–4]. The map of the retrieved $\Delta \phi (x,y)$ provides a differential phase contrast image of the sample. Furthermore, it is desirable to find out the functional relationship between the intensity fringe phase shift $\Delta \phi$ and the sample phase gradient $\partial \Phi _{\mathrm {s}}(x,y)/\partial x$ for two reasons. First, this functional relationship will enable one to retrieve a map of $\partial \Phi _{\mathrm {s}}(x,y)/\partial x$ from the measured map of the fringe phase shifts $\Delta \phi$. The retrieved sample gradients $\partial \Phi _{\mathrm {s}}(x,y)/\partial x$ from the angular projections can be used to reconstruct the 3D maps of the sample electron densities. Second, for a given projection, this functional relationship can be used to determine the ratio of $\left \vert {\Delta \phi /2\pi \alpha }\right \vert$, which is called as the angular sensitivity of a interferometry setup in literature. The angular sensitivity of a setup tells the fringe phase shift generated by a unit refraction angle. This ratio depends not only on the configuration of the interferometer and phase grating periods, but also on the position of the sample in the ray path. Hence, the angular sensitivity provides a design optimization tool for interferometry.The functional relationship between the fringe shift $\Delta \phi$ and the sample phase gradient $\partial \Phi _{\mathrm {s}}(x,y)/\partial x$ in the Talbot-Lau X-ray interferometry is well described in literature [23]. Recently we derived the corresponding functional relationship in dual phase grating X-ray interferometry [21]. In this work we set out to derive the formulas relating the fringe shift $\Delta \phi$ and the sample phase gradient $\partial \Phi _{\mathrm {s}}(x,y)/\partial x$ for triple phase grating interferometry. We hope this work will stimulate more studies on triple phase grating interferometry.
2. Methods
In order to find out the functional relationship between the fringe shift $\Delta \phi$ and the sample phase gradient $\partial \Phi _{\mathrm {s}}(x,y)/\partial x$ in triple phase grating interferometry, one should understand how the intensity fringe is formed. To start, we first give an intuitive explanation of the intensity fringe formation mechanism in the triple-phase grating interferometry. Figure 1 shows the geometrical configuration of a triple-phase grating interferometer. It consists an X-ray source $S$, three phase gratings $G_1$, $G_2$, and $G_3$ and an hospital-grade detector $D$. The periods of the three phase gratings are $p_{_{1}}$, $p_{_{2}}$, and $p_{_{3}}$ respectively. As in the geometric configuration setup of the triple-phase grating interferometry(Fig. 1), $R_s$ denotes the source-to-$G_1$ grating distance, $R_{g_{12}}$, $R_{g_{23}}$ are the spacings of gratings $G_1$, $G_2$, and gratings $G_{2}$, $G_3$ respectively, and $R_d$ marks the $G_3$-to-detector distance. For convenience, we also denote $L_i$ the source-to-$G_i$ distance, $i=1,2,3$, and $L_{\mathrm {D}}$ the source-to-detector distance. That is $L_1=R_s$, $L_2=L_1+R_{g_{12}}$, $L_3=L_2+R_{g_{23}}$, and $L_{\mathrm {D}}=L_3+R_d$. The magnification factor of $G_i$-to-detector plane is $M_{g_i}= L_{\mathrm {D}}/L_i$, $i=1,2,3$. In addition, we denote the source-to-sample distance by $L_s$.
In a previous work, based on the Wigner distribution formalism of wave diffraction, we developed a general theory of dual phase grating interferometry [19]. Using the same approach, we established a general theory for triple-phase grating interferometry. The derivation is outlined in the Appendix. According to this theory, the X-ray irradiance at the detector entrance is a sum of different diffracted orders, each of them is represented by an irradiance fringe pattern of the form $\exp \left [i2\pi x\cdot \left (l/(M_{g_1}p_{_{1}}) + r/(M_{g_2}p_{_{2}}) + v/(M_{g_3}p_{_{3}})\right )\right ]$, where $l$, $r$, and $v$ are integers. Here, without going deeply into the theoretical derivation, we offer a heuristic physics picture of the fringe formation. First imagine what is the irradiance pattern in absence of the $G_2$ an $G_3$ gratings. Obviously, the irradiance pattern would be a sum of different diffraction orders generated by the $G_1$ phase grating alone. Each of the diffraction orders would be represented by $\exp \left [i2\pi x\cdot l/(M_{g_1}p_{_{1}})\right ]$, where $l$ is an integer indexing the diffraction order. Similarly, in absence of the $G_1$ and $G_3$ gratings, the intensity pattern generated by the $G_2$ grating alone would be the sum of different diffraction orders, each of which is represented by $\exp \left [i2\pi x\cdot r/(M_{g_2}p_{_{2}})\right ]$, where $r$ indexes the diffraction order generated by the $G_2$ phase grating alone. By the same argument, the $G_3$ grating alone would generate intensity patterns comprising diffracted orders of the form $\exp \left [i2\pi x\cdot v/(M_{g_3}p_{_{3}})\right ]$. As is rigorously shown in Appendix, the X-ray irradiance at the detector entrance is a result of cross-modulation between the fringe patterns generated by each of the three phase gratings $G_1$, $G_2$ and $G_3$, respectively. As a result the irradiance pattern at the detector entrance is a weighted sum of different diffracted orders, each of which is represented by a product:
As is mentioned in the Introduction, a necessary condition for high-sensitivity interferometry is to use fine period phase gratings of micrometer or sub-micrometers. Nevertheless, in the context of grating based phase contrast imaging, it is desirable to use a hospital-grade imaging detector with pixels of tens micrometers. Hence, as long as the detector pixel size $p_{\mathrm {D}} \gg p_{1}, p_{2}, p_{3}$, most of the fine fringe patterns are rendered to a constant background. However, with proper setup, some of the fringe patterns can have much broad fringe periods and these broad fringe patterns can then be resolved by the detector. These broad fringe patterns are indeed the beat patterns corresponding to diffraction orders with a specific order, say $(l, r, v)$. To determine which $(l,r,v)$-order could correspond to broad fringes, one needs to investigate Eq. (3) to find a suitable geometric setup, i.e. appropriate $M_{g_1}$, $M_{g_2}$, $M_{g_3}$, and the order $(l,r,v)$. In Section 3.2, we give an example on how to find these broad period fringe patterns. Once we have understood the fringe formation mechanism, we are ready to tackle the problem of fringe phase shift caused by sample refraction.
Before investigating the fringe phase shift caused by sample refraction in triple-phase grating interferometry, let’s consider a single-phase grating interferometer setup as is shown in Figs. 2(a) and (b). As can be seen from Fig. 2(a), when sample is down stream from the phase grating $G$, the refracted ray causes a lateral fringe shift of $\Delta x = \alpha \cdot (L_d-L_s)$, where $L_d$ is the source-to-detector distance, $L_s$ is the source-to-sample distance. Since the period of the $l$-th order intensity fringe is $M_gp/l$, where $M_g=L_d/L_g$ is the magnification factor from grating to detector, and $p$ is the grating period, and a lateral fringe shift of one period is equivalent to a fringe phase shift of $2\pi$. Therefore, for the $l$-th order fringe, the phase shift caused by the sample refraction is:
3. Results
3.1 Fringe phase shifts caused by sample refraction
Now consider a triple-phase grating interferometer as is shown in Fig. 1, in which the gratings are marked as $G_i$ with $i = 1, 2, 3$ and various interval lengths are indicated. As is explained in Section 2, the interference fringes in triple-phase grating interferometry are consist of various diffraction orders. As is shown in Eq. (2), each of the orders is indexed by three integers $(l,r,v)$ and represented by a product of three exponentials, which represent the intensity modulation generated by each individual phase grating. This being so, the sample-generated phase shift for a fringe of order $(l,r,v)$ will be a sum of the three fringe phase shifts associated with each individual phase grating. Therefore, we have the sample-generated fringe phase shift as:
To the best of our knowledge, this is the first study that derived out the mathematical relationship between the interference fringe shift and the sample phase gradient for triple phase grating interferometry. Equation (7) shows that the sample generated fringe phase shift is proportional to the sample phase gradient with a proportion constant $\xi (l,r,v)$. Note that $\xi$ bears rich physics with it. First, the ratio $\left \vert {\xi }\right \vert /\lambda$ is equal to the angular sensitivity $\left \vert {\Delta \phi /(2\pi \alpha )}\right \vert$[21,23], which is a measure of fringe phase shift generated per unit sample-refraction angle. Hence Eq. (7) can be used as a guiding tool for optimizing angular sensitivity. Obviously, Eq. (7) dictates that $\xi =0$ when sample is placed on source plane or detector entrance. Moreover, Eq. (7) implies that for a given geometric configuration, $\xi$ is a piecewise linear function of the source-to-sample distance $L_s$ with breakpoints $L_s=0$, $L_1$, $L_2$, $L_3$, and $L_{\mathrm {D}}$. So when sample position changes within two adjacent breakpoints the angular sensitivity varies monotonically. Therefore, the maximum of $\left \vert {\xi }\right \vert$ will be achieved only at one of the breakpoints. In other words, the maximum of $\left \vert {\xi }\right \vert$ will occur at one of the grating planes. This feature revealed by Eq. (7) suggests a practical way to optimize angular sensitivity. In order to find the maximal value of $\left \vert {\xi }\right \vert$, using Eq. (7), one only needs to compare the calculated sensitivity values for sample being placed at each of the three phase grating planes and selects the largest one. Second, the proportional constant $\xi$ is also called the auto-correlation length in literature [4,5,18]. This quantity $\xi$ determines the sensitivity of a setup for detecting small angle X-ray scattering from the fine structures of the sample. But this topic is out of the scope of this study [4,5,18].
To see how to use the general formula of Eq. (7), we divided possible sample positions into four intervals. They are (i) the interval between source and $G_1$ grating, (ii) that between $G_1$ and $G_2$ gratings, (iii) that between $G_2$ and $G_3$ grating, and finally, (iv) that between $G_3$ grating and the detector entrance. We applied Eq. (7) in each of the four intervals and found the corresponding interference fringe phase shifts as follows.
In interval (i) the fringe shift is given by:
3.2 Predictions on fringe period and angular sensitivity of a triple-phase grating interferometry setup
The performance of a triple-phase grating interferometer is governed by several aspects: (i) if the generated fringe period is large enough to be detected by a common detector; (ii) if the visibility of the fringe pattern is high enough to tolerate noise; (iii) with a given triple-phase grating interferometer, where should we place the sample to get higher angular sensitivity?
Specifically, assume we employ three $\pi$-phase gratings of design energy ${E_{\mathrm {D}}}=20$keV, and X-ray beam is monochromatic [17]. The periods for the three gratings are $p_{_{1}}=p_{_{3}}=2p_{_{2}}=p=1\mu$m. According to Eq. (3) in Section 2, the fringe period for diffracted order $(l, r, v)$ is
To affirm the assertion above, we perform numerical simulation through wave propagation. In the simulation we assume a $20$keV point X-ray source and a detector of period $p_{\mathrm {D}}=20\mu$m. The three phase gratings are $\pi$ gratings at design energy of $20$keV and periods $p_{_{1}}=p_{_{3}}=2p_{_{2}}=1\mu$m [24]. As is stated in the last paragraph, with the setup of $L_{\mathrm {D}}=2$m, and $(R_s, R_{g_{12}}, R_{g_{23}}, R_d)=(0.5, 0.5, 0.51, 0.49)$ meters, the beating fringe period is ${p_{\mathrm {fr}}}=100\mu$m. So one fringe period will occupies $5$ pitches in the detector. Figure 3 shows the intensity fringe pattern, which confirms the theoretical results. Several fringe patterns of different X-ray energies are shown in Fig. 3. As photon energy changes, the fringe diffraction order may either be unchanged, or be changed to the $l=1$ diffraction order from $l=2$ for the triple $\pi$ grating setups. Hence, the fringe period is either unchanged or is doubled, as is shown by Eq. (12), or by the general formula of Eq. (3). Meanwhile, the magnitude of fringe modulation varies significantly with photon energy. Under this setup, the fringe visibility, defined as $V=(\max (I)-\min (I))/(\max (I)+\min (I))$, is as high as $V=0.46$ as is shown from numerical simulation in Fig. 3 for X-ray energy ${E_{\mathrm {D}}}=20$ keV.
Now with the given setup, we turn on to find the optimal sample position so that the angular sensitivity is maximized. First, let us look at what Eq. (7) predicts. Recall from Eq. (7) that the factor $\xi$ is a piecewise linear function of sample position $L_s$. $\left \vert {\xi }\right \vert$ is maximized when sample is at one of the grating plane, since $\xi =0$, when $L_s=0$ or $L_{\mathrm {D}}$. By setting $L_s=L_1$, $L_2$, or $L_3$, one finds the corresponding values of $\xi$ are $-(5.00\mathrm {E}3)\lambda$, $-(1.01\mathrm {E}6)\lambda$, and $(4.90\mathrm {E}3)\lambda$ respectively. Here $\lambda$ is the X-ray wavelength, which equals to $6.2\mathrm {E}\!-\!\!11$m for a $20$keV X-ray source. A plot of $\xi /\lambda$ with respect to $L_s$ is shown in Fig. 4, where $L_s$ spans from $L_1/2$ to $(L_3+ L_{\mathrm {D}})/2$. Thus Eq. (7) predicts that $L_s=L_2$ is an optimal sample position where the interferometer’s angular sensitivity is maximized.
3.3 Simulations to validate the theory
To validate the Eq. (7) in Section 3, we perform a numerical simulation with a sphere of diameter $5$mm placed half way between the second and third gratings with $L_s=1.255$m. The sphere is filled with a hypothetical tissue, which has the same chemical composition as breast adipose tissue but with lower mass density. In this way the retrieved fringe phase shifts are less than $\pi$. We employed ray-tracing to get the projected sample attenuation $A^{2}(x,y)$, and the phase map of the sample. From the sample phase map we computed the sample phase gradients. Using Eq. (7) and the sample phase gradients we compute the theoretical map of the fringe shift $\Delta \phi (x,y)$ caused by sample refraction. The theoretical map of the sample attenuation $A^{2}(x,y)$ and the map of the fringe phase shift $\Delta \phi (x,y)$ are shown in Figs. 6(a) and 6(b) respectively. We then employed Fresnel propagation down stream through $G_1$, $G_2$, the sample sphere, the grating $G_3$ and finally to the detector plane to get the projected image. The simulated fringe patterns at the detector, with and without samples, are shown in Fig. 5. The fringe phase shift $\Delta \phi (x, y)$ and the attenuation map $A^{2}(x, y)$ were retrieved with the Fourier fringe analysis method developed in [14–16,25], which we call the FT-retrieval for short. The maps of the retrieved attenuation and the fringe phase shift are shown in Figs. 6(b) and 6(e) respectively. As comparison, profiles of the attenuation $A^{2}(x)$ and the fringe phase shift $\Delta \phi (x)$ along the central line across grating are shown in Figs. 6(c) and 6(f) respectively. The red-dash curves represent the theoretical values, while the blue-solid curves are the retrieved values from the intensity fringes. The good match between the theoretical and FT-retrieved values of $\Delta \phi (x)$ validates the fringe phase shift formula of Eq. (7). The The oscillations in Fig. 6(c) is due to the FT-phase retrieval errors, which results from the rapidly varying attenuation at the sphere boundary. To avoid the infinity values of the phase gradient at the edges of the sphere, in the simulation we slightly modified the sphere boundary to render the boundary values finite.
4. Discussion and conclusions
The functional relationship between fringe phase shifts and sample phase gradients, as is derived in Eq. (7), enables the retrieval of sample phase gradient in triple phase grating interferometry. Obviously, it lays a foundation for quantitative differential phase contrast imaging. Moreover, the volumetric images of the sample electron densities can be reconstructed from the sample gradients retrieved in different angular projections. On the other hand, for implementing triple phase grating interferometry, Eq. (7) is especially useful. In fact, the angular sensitivity of an interferometry setup is an indicator of its imaging sensitivity. The angular sensitivity of an interferometer depends not only on the phase shifts and periods of the three phase gratings, but also on the geometric configuration of the interferometer and the sample position. There is a pressing need of a design optimization tool to predict how sensitive a specific setup is. The functional relationship derived in Eq. (7) enables one to predict the angular sensitivity of an interferometer setup with triple phase gratings. In this work, as an example, we studied the angular sensitivities of three $\pi$ phase grating setups. As is shown in Fig. 4, this interferometer demonstrates large variations in the angular sensitivity for different sample placements. Hence Eq. (7) provides a useful tool for system optimization. We stress that, our formula of Eq. (7) is applicable to any triple phase grating setups. For example, one may adopt setups with $\pi /2$-$\pi$-$\pi /2$ phase gratings [17]. By using Eq. (9), we found that, in the $\pi /2$-$\pi$-$\pi /2$ setups with identical grating period of $1\mu$m, the angular sensitivity peaks as well when the sample is on the $G_2$ grating plane, but the sensitivity is different. This is because the beat fringe patterns of the three $\pi$ grating setups attain a diffraction orders $(l,r,v)=(2,-2,2)$ and its complex conjugate, but the beat fringe patterns of $\pi /2$-$\pi$-$\pi /2$ setups are of diffraction orders $(l,r,v)=(1,-2,1)$ and its complex conjugate. So Eq. (7) dictates that the angular sensitivity of the $\pi /2$-$\pi$-$\pi /2$ setups are half that of the three $\pi$ phase grating setups. Due to the space limitation, we are unable to discuss more details about the various implementations of triple phase grating interferometry.
Compared to Talbot-Lau interferometry, a decisive advantage of triple phase grating setups with fine grating periods lies in that the fringe period can be directly resolved by common imaging detectors. This is good for medical imaging applications, as is explained in the introduction. Of course, the dual phase grating interferometry can be implemented with common imaging detectors as well. As for angular sensitivities with dual phase grating setups, we derived a corresponding formula in a previous work [18,21]. In that work we showed that, for dual phase grating setups, the optimal sample placement is on the first grating plane. Apparently, this conclusion is different from what we found in Section 3 for setups with triple phase gratings. Compared on the basis of the same grating periods and total system length, an optimal triple phase grating setup can achieve few times higher in angular sensitivity than that with dual phase grating setups. With triple phase grating setups, there are more pathways for wave interference and beat patterns formation, as compared to the dual phase grating setups. In addition, in dual phase grating interferometry the spacing between the two phase gratings are very narrow for allowing large fringe period [19]. In the triple phase grating interferometry the space between gratings are usually ample and convenient for sample placement. In addition to fringe visibility, the average intensity of the fringe patterns also determines the quality of the sample phase gradient images. Exactly speaking, it is the ratio of the 1st-order to the 0th-order Fourier components of the fringe patterns that determines image quality [26]. As a disadvantage of triple phase grating interferometry, note that its fringe visibility is generally lower than that of the Talbot-Lau and the dual phase grating interferometry. This is because the intensity modulation is essentially a product of the modulation contributed by each of the three phase gratings, as is shown in Appendix. Another limitation of this work lies in that we did not include polychromatic X-ray in our derivation. Polychromatic X-ray can distort fringe phase shifts by spectral averaging and beam hardening [27–30]. More research is needed on these issues.
In conclusion, in this work, by using intuitive analysis of the beat pattern formation and sample refraction in triple phase grating interferometry, the authors derived exact formulas relating sample phase gradients to fringe phase shifts. These formulas not only provide a design optimization tool for triple phase grating interferometry, but also lay a foundation for quantitative phase contrast imaging.
NB In the manuscript review process, we were informed by a reviewer that there is a recent preprint [31], which also investigates the angular sensitivity of triple phase grating interferometry. Different from our work, the core method adopted by this preprint is based on the thin lens imaging theory developed for the dual phase grating interferometer [22]. The preprint does not present any explicit formula of angular sensitivity for triple grating interferometry.
Appendix
In this Appendix, we prove the following assertions: For the interferometry shown in Fig. 1, assume the phase gratings $G_i$, $i=1,2,3$ are binary phase gratings with duty cycles of 0.5. The phase shifts and periods of the phase gratings $G_i$ are denoted by $\phi _i$ and $p_{_{i}}$ respectively, $i=1,2,3$. Let $L_i$ be the source-to-$G_i$ grating distance, $i=1,2,3$, and $L_{\mathrm {D}}$ be the source-to-detector distance. Then the X-ray irradiance $I(x,y)$ at the detector entrance can be written in the form
Due to the space limitation, we present an outline of the proof. We use the Wigner function formalism to derive a general form of the equation. The Wigner function of an X-ray wavefront at $z=s_1$ is defined as the Fourier transform of the mutual intensity of the wavefront $J_{s_1}(x_1, x_2;y)$[32–34]:
Since $S_i$ is a periodic function of period $p_{_{i}}$, $i=1,2,3$ (see Eqs. (23), (30), and (32)), following the same arguments presented in the Appendix of [35], one can show that for real numbers $v$ and $s$,
where $\delta$ is the Dirac delta function, and when $v=n/p_{_{i}}$, $n$ is integer, $\Psi _i$ can be analytically expressed in the form of Eq. (18). Substitute Eq. (36) back to Eq. (34), after a tedious computation, one finishes the proof of the assertion.Funding
National Institutes of Health (1R01CA193378).
Disclosures
The authors declare no conflicts of interest.
References
1. A. Momose, S. Kawamoto, I. Koyama, Y. Hamaishi, H. Takai, and Y. Suzuki, “Demonstration of X-ray talbot interferometry,” Jpn. J. Appl. Phys. 42(Part 2, No. 7B), L866–L868 (2003). [CrossRef]
2. T. Weitkamp, A. Diaz, C. David, F. Pfeiffer, M. Stampanoni, P. Cloetens, and E. Ziegler, “X–ray phase imaging with a grating interferometer,” Opt. Express 13(16), 6296–6304 (2005). [CrossRef]
3. A. Momose, “Recent advances in X-ray phase imaging,” Jpn. J. Appl. Phys. 44(9A), 6355–6367 (2005). [CrossRef]
4. F. Pfeiffer, T. Weitkamp, O. Bunk, and C. David, “Phase retrieval and differential phase-contrast imaging with low-brilliance X-ray sources,” Nat. Phys. 2(4), 258–261 (2006). [CrossRef]
5. W. Yashiro, Y. Terui, K. Kawabata, and A. Momose, “On the origin of visibility contrast in X-ray talbot interferometry,” Opt. Express 18(16), 16890–16901 (2010). [CrossRef]
6. P. Zhu, K. Zhang, Z. Wang, Y. Liu, X. Liu, Z. Wu, S. McDonald, F. Marone, and M. Stampanoni, “Low-dose, simple, and fast grating-based X-ray phase-contrast imaging,” Proc. Natl. Acad. Sci. U. S. A. 107(31), 13576–13581 (2010). [CrossRef]
7. N. Bevins, J. Zambelli, K. Li, Z. Qi, and G.-H. Chen, “Multicontrast X-ray computed tomography imaging using talbot-lau interferometry without phase stepping,” Med. Phys. 39(1), 424–428 (2011). [CrossRef]
8. X. Tang, Y. Yang, and S. Tang, “Characterization of imaging performance in differential phase contrast ct compared with the conventional ct: Spectrum of noise equivalent quanta neq(k),” Med. Phys. 39(7Part1), 4467–4482 (2012). [CrossRef]
9. T. Teshima, Y. Setomoto, and T. Den, “Two-dimensional gratings-based phase-contrast imaging using a conventional X-ray tube,” Opt. Lett. 36(18), 3551–3553 (2011). [CrossRef]
10. E. Bennett, R. Kopace, A. Stein, and H. Wen, “A grating-based single shot X-ray phase contrast and diffraction method for in vivo imaging,” Med. Phys. 37(11), 6047–6054 (2010). [CrossRef]
11. M. Jiang, C. Wyatt, and G. Wang, “X-ray phase-contrast imaging with three 2d gratings,” Int. J. Biomed. Imaging 2008, 1–8 (2008). [CrossRef]
12. H. Itoh, K. Nagai, G. Sato, K. Yamaguchi, T. Nakamura, T. Kondoh, C. Ouchi, T. Teshima, Y. Setomoto, and T. Den, “Two-dimensional grating-based X-ray phase-contrast imaging using fourier transform phase retrieval,” Opt. Express 19(4), 3339–3346 (2011). [CrossRef]
13. A. Bravin, P. Coan, and P. Suortti, “X-ray phase-contrast imaging: from pre-clinical applications towards clinics,” Phys. Med. Biol. 58(1), R1–R35 (2013). [CrossRef]
14. A. Momose, H. Kuwabara, and W. Yashiro, “X-ray phase imaging using lau effects,” Appl. Phys. Express 4(6), 066603 (2011). [CrossRef]
15. N. Morimoto, S. Fujino, K. Ohshima, J. Harada, T. Hosoi, H. Watanabe, and T. Shimura, “X-ray phase contrast imaging by compact talbot-lau interferometer with a single transmission grating,” Opt. Lett. 39(15), 4297–4300 (2014). [CrossRef]
16. N. Morimoto, S. Fujino, A. Yamazaki, Y. Ito, T. Hosoi, H. Watanabe, and T. Shimura, “Two dimensional X-ray phase imaging using single grating interferometer with embedded X-ray targets,” Opt. Express 23(13), 16582–16588 (2015). [CrossRef]
17. H. Miao, A. Panna, A. Gomella, E. Bennett, S. Znati, L. Chen, and H. Wen, “A universal moiré effect and application in X-ray phase-contrast imaging,” Nat. Phys. 12(9), 830–834 (2016). [CrossRef]
18. M. Kagias, Z. Wang, K. Jefimovs, and M. Stampanoni, “Dual phase grating interferometer for tunable dark-field sensitivity,” Appl. Phys. Lett. 110(1), 014105 (2017). [CrossRef]
19. A. Yan, X. Wu, and H. Liu, “Quantitative theory of X-ray interferometers based on dual phase grating: fringe period and visibility,” Opt. Express 26(18), 23142–23155 (2018). [CrossRef]
20. A. Yan, X. Wu, and H. Liu, “Clarification on generalized lau condition for X-ray interferometers based on dual phase gratings,” Opt. Express 27(16), 22727–22736 (2019). [CrossRef]
21. A. Yan, X. Wu, and H. Liu, “Sample phase gradient and fringe phase shift in dual phase grating X-ray interferometry,” Opt. Express 27(24), 35437–35447 (2019). [CrossRef]
22. Y. Ge, J. Chen, P. Zhu, J. Yang, S. Deng, W. Shi, K. Zhang, J. Guo, H. Zhang, H. Zheng, and D. Liang, “Dual phase grating based X-ray differential phase contrast imaging with source grating: theory and validation,” Opt. Express 28(7), 9786–9801 (2020). [CrossRef]
23. T. Donath, M. Chabior, F. Pfeiffer, O. Bunk, E. Reznikova, J. Mohr, E. Hempel, S. Popescu, M. Hoheisel, M. Schuster, J. Baumann, and C. David, “Inverse geometry for grating-based X-ray phase-contrast imaging,” J. Appl. Phys. 106(5), 054703 (2009). [CrossRef]
24. H. Miao, A. A. Gomella, K. J. Harmon, E. E. Bennett, N. Chedid, A. Panna, P. Bhandarkar, and H. Wen, “Spherical-wave far-field interferometer for hard X-ray phase contrast imaging,” arXiv:1412.6824 (2014).
25. M. Takeda, H. Ina, and S. Kobayashi, “Fourier-transform method of fringe-pattern analysis for computer-based topography and interferometry,” J. Opt. Soc. Am. 72(1), 156–160 (1982). [CrossRef]
26. W. Yashiro, Y. Takeda, and A. Momose, “Efficiency of capturing a phase image using cone-beam X-ray talbot interferometry,” J. Opt. Soc. Am. A 25(8), 2025–2039 (2008). [CrossRef]
27. A. Yan, X. Wu, and H. Liu, “Polychromatic X-ray effects on fringe phase shifts in grating interferometry,” Opt. Express 25(6), 6053–6068 (2017). [CrossRef]
28. A. Yan, X. Wu, and H. Liu, “Beam hardening correction in polychromatic X-ray grating interferometry,” Opt. Express 25(20), 24690–24704 (2017). [CrossRef]
29. A. Hipp, M. Willner, J. Herzen, S. Auweter, M. Chabior, J. Meiser, K. Achterhold, J. Mohr, and F. Pfeiffer, “Energy-resolved visibility analysis of grating interferometers operated at polychromatic X-ray sources,” Opt. Express 22(25), 30394–30409 (2014). [CrossRef]
30. A. Pandeshwar, M. Kagias, Z. Wang, and M. Stampanoni, “Modeling of beam hardening effects in a dual-phase X-ray grating interferometer for quantitative dark-field imaging,” Opt. Express 28(13), 19187–19204 (2020). [CrossRef]
31. J. Chen, J. Yang, P. Zhu, T. Su, H. Zhang, H. Zheng, D. Liang, and Y. Ge, “Estimation of angular sensitivity for X-ray interferometers with multiple phase gratings,” arXiv:2006.13351v1 (2020).
32. L. Mandel and E. Wolf, Optical Coherence and Quantum Optics (Cambridge University, 1995).
33. X. Wu and H. Liu, “A new theory of phase-contrast X-ray imaging based on wigner distributions,” Med. Phys. 31(9), 2378–2384 (2004). [CrossRef]
34. X. Wu and H. Liu, “Phase-space evolution of X-ray coherence in phase-sensitive imaging,” Appl. Opt. 47(22), E44–52 (2008). [CrossRef]
35. A. Yan, X. Wu, and H. Liu, “A general theory of interference fringes in X-ray phase grating imaging,” Med. Phys. 42(6Part1), 3036–3047 (2015). [CrossRef]