Abstract
We show that a linear relation exists between the device sensitivity and the quality (Q) factor of a dual-waveguide coupled microring resonator optical biosensor when the optimal conditions are satisfied. We also show that the detection limit depends on the loss coefficient and signal-to-nosie ratio (SNR) of the overall system, rather than the circumference of the ring. For a microring resonator sensor whose Q factor is 20000, the detection limit is found to be about 10^{-7} with 30-dB SNR, which is in good agreement with reported experimental data. These results indicate that loss reduction is the top priority in the design and fabrication of highly sensitive microring resonator optical biosensors.
© 2009 Chinese Optics Letters
PDF Article
More Like This
Cited By
You do not have subscription access to this journal. Cited by links are available to subscribers only. You may subscribe either as an Optica member, or as an authorized user of your institution.
Contact your librarian or system administrator
or