Abstract
In a previously published work, a procedure was described to conduct qualitative analysis in inductively coupled plasma atomic emission spectrometry, keeping in mind the possibility of spectral interferences. This procedure is based on the use of both line correlation and normalized relative line intensities of given elements. When spectral interferences due to a single or two major elements are observed for an analyte, use of multiple linear regression of the normalized relative line intensities of both the analyte and the major element provides information about the certainty of the presence of the analyte. This procedure has been used with an echelle grating-based dispersive system equipped with a custom segmented-array charge-coupled device detector and evaluated with the use of line-rich matrices such as Fe, Cr, and Ni. Satisfactory results were obtained even when several lines of the analyte suffered from interference. The only limitation was the number of available lines for a given analyte.
PDF Article
More Like This
Multielemental analysis of prehistoric animal teeth by laser-induced breakdown spectroscopy and laser ablation inductively coupled plasma mass spectrometry
Michaela Galiová, Jozef Kaiser, Francisco J. Fortes, Karel Novotný, Radomír Malina, Lubomír Prokeš, Aleš Hrdlička, Tomáš Vaculovič, Miriam Nývltová Fišáková, Jiří Svoboda, Viktor Kanický, and Javier J. Laserna
Appl. Opt. 49(13) C191-C199 (2010)
Cited By
You do not have subscription access to this journal. Cited by links are available to subscribers only. You may subscribe either as an Optica member, or as an authorized user of your institution.
Contact your librarian or system administrator
or