Abstract
Results are presented on the statistical fluctuations occurring in a forward-light-scattering experiment to determine the particle size distribution. A sample of glass beads was measured using a Malvern 2600D instrument and analyzed with a proposed deconvolution procedure that incorporates the observed intensity fluctuations. This procedure yields a qualitative improvement of the solution, provides error intervals, and offers a better means for model discrimination.
© 1991 Optical Society of America
Full Article | PDF ArticleMore Like This
Karl A. Kusters, Johan G. Wijers, and Dirk Thoenes
Appl. Opt. 30(33) 4839-4847 (1991)
Gorden Videen, Paul Pellegrino, Dat Ngo, John S. Videen, and Ronald G. Pinnick
Appl. Opt. 36(24) 6115-6118 (1997)
Douglas A. Ross
Appl. Opt. 30(33) 4882-4888 (1991)